DLP(R) Pericardial/Intracardiac Sump Cannulae, 20 Fr., P/N: 12112. This product is intended for draining the pericardial sac or the cardiac chambers.

FDA Recall Z-0609-2009 · Posted 2009-01-26

Recall StatusTerminated
Product CodeDWF
Date Initiated2008-09-29
Date Terminated2011-05-07
Product Quantity380 US ; 180 OUS

Reason for Recall

Separates: Six complaints were received from two locations. In all cases the sump tip assembly separated from the tubing during device removal and had to be retrieved from the patient. There were no adverse events affecting these patients.

Firm Action

Consignees were sent a "Medtronic Urgent Medical Device Recall Notice" dated October 1, 2008. The letter described the product and issue. It also requested its customers to return the product and to complete, sign and fax the Recall Certificate "Medtronic Recall Certificate (FCA 0902)". For further information, please contact Medtronic Cardiovascular Revascularization & Surgical Therap by telephone at 763-391-9000.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution --- including states of NC, PA, IA, WI, IN, VA, IL, KY, and TX, and countries of Australia, France, Hong Kong, Netherlands, and Poland.