Datex-Ohmeda Cardiocap/5 Anesthesia Monitor with N-XOSAT SpO2 measurement option

FDA Recall Z-0611-03 · Posted 2003-03-05

Recalling FirmDatex-Ohmeda
Recall StatusTerminated
Product CodeDPZ
Date Initiated2002-09-16
Date Terminated2004-07-02
Product Quantity85

Reason for Recall

Falsely high blood oxygen saturation readings in some circumstances which could result in no alarm in a critical life-threatening situation.

Firm Action

'Medical Device Advisory Notice' letters dated September 16, 2002 were sent to the consignees, gave recommendations about alternative monitoring for certain patients, and stated that there would be a field correction to reduce the possibility of the problem.

Root Cause

Other

Distribution Pattern

Nationwide in the United States and worldwide.