Chlamydia OIA. For in vitro diagnostic use,
FDA Recall Z-0615-03 · Posted 2003-03-06
Advertisement
Recalling FirmThermo Biostar, Incorporated
Recall StatusTerminated
Product CodeLJC
Date Initiated2003-02-07
Date Terminated2003-05-02
Product Quantity286 units
Reason for Recall
Chlamydia IVD test kits may contain diluted conjugate reagent. Use may produce a false negative result with a low positive patient sample.
Firm Action
All consignees were notified by telephone on 2/6/2003, followed by letter on 2/10/2003.
Root Cause
Other
Distribution Pattern
Nationwide, and to France, New Zealand and Spain.