Terminated

Abbott i-STAT CG8+ Cartridges intended for use in the in vitro quantification of potassium in arterial, venous, or capillary whole blood. List Number/UDI# : 03P88-25 ,(01)10054749000163(17)181028(10)L18059

FDA Recall Z-0615-2019 · Posted None

Recall StatusTerminated
Product CodeCHL
Date Initiated2018-09-13
Date Terminated2019-04-04
Product Quantity17550 cartridges

Reason for Recall

Inaccurate potassium results may be generated. When testing with control fluid and/or patient samples with low or reference range potassium concentration, elevated potassium results may be generated. When testing with samples with a high concentration of potassium, decreased potassium results may be generated

Firm Action

Abbott Point of Care issued recall letter on 9/13/18 via FedEx , advising of the problem, recommend action to return all unused cartridges to APOC . If you have any questions regarding this information, please contact Abbott Point of Care Technical Support at 844-256-9531 or your Abbott Point of Care representative.

Root Cause

Manufacturing material removal

Distribution Pattern

WI