Terminated

DeVilbiss Model 515A Oxygen Concentrator Intended to provide supplemental low flow oxygen therapy. These devices can be used in homes, nursing homes, and patient care facilities.

FDA Recall Z-0617-2009 · Posted 2008-12-30

Recalling FirmSunrise Medical
Recall StatusTerminated
Product CodeCAW
Date Initiated2008-10-30
Date Terminated2012-10-27
Product Quantity428 units

Reason for Recall

potential for sieve beds to have been incorrectly assembled

Firm Action

On 10/30/08, the recalling firm began to telephone customers of the problem. The recalling firm followed up with Urgent: Medical Device Recall letters, dated 11/7/2008. The letters requested that users cease the use of the affected device and immediately contact their provider to make arrangements to have the Concentrator replaced. The firm has set up a toll free line to assist customers with identification of affected units and the return process.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution --- USA including states of AL, CA, FL ,GA, IN, IL, MA, MD, MI, MS, NC, NE, NH, NM, NY, PA, SD, TN, TX, VA, and WI, and countries of Canada, Columbia, Germany, Mexico, Netherlands, Singapore, and United Kingdom.