Flexi-Seal Fecal Management System Advanced Odor Control Kit. Rx only; 1 silicone catheter tube assembly, 3-1 Liter collection bags with filters, 1 luer-lock syringe, SKU #411104 ConvaTec Skillman, NJ Made in USA with imported components "For the fecal management of patients with little or no bowel control and liquid or semi-liquid stool."
FDA Recall Z-0618-2009 · Posted 2008-12-30
Advertisement
Recalling FirmConvaTec
Recall StatusTerminated
Product CodeKNT
Date Initiated2008-11-20
Date Terminated2009-03-09
Product Quantity795
Reason for Recall
It was discovered that a portion of the inner nylon sleeve of the catheter was visible at the balloon end of a sample device.
Firm Action
Urgent Product Correction Letters were sent by first class mail on November 18, 2008 to Healthcare Professionals and ConvaTec Reps. The reason for the recall was identified along with possible risk to patient. Healthcare professionals were asked to use sufficient lubricating jelly during the insertion process with special care given to those who have a tendency to bleed or are on anti-coagulant medication. Physician should be notified if rectal bleeding or abdominal symptoms develop. ConvaTec reps were asked to contact all health care professionals to whom product or samples were given and reinforce the importance of proper lubrication and following the directions. A return form is to be completed by ConvaTec reps for each visit indicating the name of the Health Care Provider and that the they personally received notice of the product correction. All remaining sample products are to be returned to Besta, Franklin, WI.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution --- including USA and country of Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement