Stryker Titanium Intraline Anchor, Model Numbers 3910400050, 3910400055, 3910400060, 3910400065, Soft Eyelet Anchor, manufactured by Stryker Endoscopy, Puerto Rico
FDA Recall Z-0618-2011 · Posted 2010-12-14
Advertisement
Recalling FirmStryker Endoscopy
Recall StatusTerminated
Product CodeHWC
Date Initiated2009-08-20
Date Terminated2010-12-14
Product Quantity447 units
Reason for Recall
Potential breakage -- product may break while inserting it into the bone during surgery. This may prolong surgery, or, if undetected during surgery, result in retained fragments that may cause metal sensitization or immune response, or necessitate further intervention to retrieve the titanium fragments.
Firm Action
Stryker issued an Urgent Device Removal notification letter dated August 21, 2009 to consignees. Consignees were asked to acknowledge receipt and distribute the letter, track reponses, and destroy any returned material and document through weekly summary and records.
Stryker can be contacted at 408 754-2000 about this recall.
Root Cause
Material/Component Contamination
Distribution Pattern
Nationwide Distribution: to one distributor in NJ.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.