Terminated

Norian Drillable Inject

FDA Recall Z-0618-2018 · Posted None

Recalling FirmSynthes USA
Recall StatusTerminated
Product CodeMQV
Date Initiated2017-08-10
Date Terminated2018-09-10
Product Quantity5 units

Reason for Recall

The contract supplier notified DePuy Synthes on June 16, 2017 that the kit expiration dates do not align with the earliest expiration date of the component lots within the kit. Some kit expiration dates were later than the earliest expiration date of the component lots. The expiration dates of some components are earlier than that listed on the kit.

Firm Action

On August 15, 2017 an Urgent Medical Device Recall notice, titled "895204 NORIAN DRILLABLE INJECT" was issued to customers instructing them to review inventory and quarantine all affected product subject to the recall. For questions. concerns or to obtain a return authorization Call 1-800-479-6329. Returned product will be credited.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution.