Inter-Op Acetabular System Hooded Trial Insert size 32mm/57mm, non-sterile, REF 9363-32-057, Zimmer, Warsaw, IN.
FDA Recall Z-0619-2011 · Posted 2010-12-14
Advertisement
Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeHWT
Date Initiated2010-11-15
Date Terminated2011-05-23
Product Quantity3
Reason for Recall
Mislabeled - Device packages labeled as hooded trial inserts contain standard trial inserts.
Firm Action
Zimmer contacted the two affected accounts directly by phone on November 15, 2010 to inform them of the recall. Zimmer will monitor the return of the affected instruments and account for the outstanding instruments until the removal action is complete. Each of the affected accounts will be provided with a new trial insert, manufactured and packaged with the correct component. The accounts will be asked to return the affected units back to Zimmer where the returned product will be destroyed.
Customers can contact Zimmer at 574 371-8852 concerning this recall.
Root Cause
Employee error
Distribution Pattern
Worldwide Distribution: US in the states of KS and UT, and the country of Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.