Foot switch, Part Number 1009.81. Maquet GmbH & Co. KG, 76437 Rastatt, Germany. The foot switches in the 1009.8 1 series are accessory components which can be connected to certain MAQUET operating tables. Several of the table top adjustment motions (depending on the table and foot switch versions in use) can then be executed by the foot. The foot switch is equipped with rocker switches. The switching elements are used to initiate the table top functions.

FDA Recall Z-0620-2009 · Posted 2009-01-15

Recalling FirmMAQUET Inc.
Recall StatusTerminated
Product CodeFWZ
Date Initiated2008-12-08
Date Terminated2012-06-18
Product Quantity587 units total (54 units in US)

Reason for Recall

A bearing point of the pedal of the foot switch may come loose. With continued use of the foot switch, one or both springs beneath the pedal can fall out of their mounting, and it is possible the pedal comes to rest on the switch causing an unintended motion of the operating table during patient use.

Firm Action

An Urgent Device Correction letter (with an enclosed Customer Response form) was sent via Federal Express on Monday, December 8, 2008 to all customers who had received tables. Customers were notified in the letter of the problem and actions to be taken including not to use any affected foot pedal for safety reasons. In addition, customers were instructed to complete and return the Customer Response Form via fax (732-667-1903) to Maquet Inc. Please call USA Maquet, Inc. at toll free number 1-888-627-8383 #1 (then select the option for Surgical Workplace products) for questions.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution.