Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190

FDA Recall Z-0620-2025 · Posted 2024-12-05

Recall StatusOpen, Classified
Product CodeFDS
Date Initiated2024-10-31
Product Quantity1,770 units

Reason for Recall

High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).

Firm Action

Olympus notified consignees via letter on about 10/31/2024. Consignees were informed of the issue and of the upcoming revision of the labeling such that the GIF-1TH190 is no longer a compatible endoscope for reprocessing with Acecide-C in the OER-Pro. They were instructed that they would be contacted by their Olympus sales representative within the month to discuss potential OER-Elite and GIF-1TH190 transition programs. Additionally, they were instructed to carefully read the notification, ensure reprocessing staff and other users of the OER-Pro AER are aware of the corrective action, acknowledge receipt of the notification through the Olympus web portal, and to identify and notify recipients of the affected devices if they were further distributed.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑