Genicon 5mm Pyramidal Trocar, Sterile, Genion 6869 Stapoint Court, Suite 114 Winter Park, FL U.S.A. 32792, Rx Only. Catalog Numbers: 100-005-001 & 100-005-002. Used for incisions made for positioning the laparoscopic cannulas or ports.

FDA Recall Z-0622-2009 · Posted 2008-12-24

Recall StatusTerminated
Product CodeGEA
Date Initiated2008-09-25
Date Terminated2009-01-09
Product Quantity2,965 units (1340 OUS and 1625 US)

Reason for Recall

The stainless steel tips on the Genicon 5mm Pyramidal Trocar exhibited contamination that resembled rust and there was also pitting on the tip of the trocar.

Firm Action

Consignees were notified with a letter, dated 9/19/2008, by e-mail and fax on 09/25/2008. The letter stated a description of the issue, analysis and reliability data, indications of Device Malfunction, and recommendations. There was also a request for all defective devices to be returned to the distributor or Genicon for replacement. Genicon Customer service was to be contacted at 407-657-4851.

Root Cause

Manufacturing material removal

Distribution Pattern

Worldwide Distribution --- including USA and countries of Africa, United Kingdom, Denmark, Colombia, Belgium, Italy, Finland, Hong Kong, and Switzerland.