UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.
FDA Recall Z-0622-2016 · Posted None
Advertisement
Recalling FirmOrthoPediatrics Corp
Recall StatusTerminated
Product CodeNKB
Date Initiated2015-11-10
Date Terminated2016-05-23
Product Quantity102
Reason for Recall
Response Spine UniAxial Pedicle & UniAxial Reduction Pedicle Screws may have been manufactured incorrectly, thus preventing the driver from fully engaging into the hexagonal drive feature of the screw.
Firm Action
The firm, OrthoPediatrics Corp., sent an "URGENT MEDICAL DEVICE RECALL" letter dated 11/13/2015 to affected customers via email. The letter included a description of the reason for the recall, affected product, consignee responsibilities, and actions to be taken. The customers were instructed to take immediate possession of all the identified devices; quarantine them and immediately return them to OrthoPediatrics, and complete and return the Mandatory Reply Form via fax to (574) 269-3692 or email: [email protected].
If you have any questions, please contact VP-Regulatory Affairs (574) 268-6379 or email: [email protected].
Root Cause
Process change control
Distribution Pattern
US Distribution to states of: AL, AZ, CA, DE, FL, GA, HI, KY, LA, MI, MO, MN, MS, NC, NJ, NY, OH, OR, TN, and TX.
Official Source