DeRoyal CARDIAC CATH SET-UP KIT PGYBK, REF 89-10572
FDA Recall Z-0622-2024 · Posted 2024-09-30
Advertisement
Recalling FirmDeRoyal Industries Inc
Recall StatusOpen, Classified
Product CodeLRO
Date Initiated2023-11-22
Product Quantity252 units
Reason for Recall
Kit were manufactured and distributed with Edwards Life Sciences TruWave Disposable pressure transducers which were subsequently recalled by Edwards Life Sciences.
Firm Action
DeRoyal Industries issued am URGENT FIELD CORRECTIVE ACTION notice to its sole consignee on 11/22/2023 via email. The notice explained the issue and provided instructions on the proper use of the device.
Root Cause
Device Design
Distribution Pattern
US: Virginia