e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203

FDA Recall Z-0623-03 · Posted 2003-03-06

Recall StatusTerminated
Product CodeKPS
Date Initiated2003-02-10
Date Terminated2004-09-15
Product Quantity1927 units

Reason for Recall

Unexpected release of the radial and rotate brakes in the case of multiple simultaneous fault conditions.

Firm Action

Siemens sent Customer Advisory letters dated 2/10/03 to all e.cam customers. The accounts were informed of the problem with unexpected camera movement when multiple fault cocnditions are presnet, causing the brake system to release and the roation of the detector assemblies, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended release of the radial and rotate brakes. The upgrade will be done free of charge.

Root Cause

Other

Distribution Pattern

Nationwide, United Arab Emirates, Argentina, Austria, Australia, Bosnia, Barbados, Bangladesh, Belgium, Bolivia, Brazil, Camaroon, Canada, Chile, China, Colombia, Republic of Congo, Czech Republic, Denmark, Dominican Republic, El Salvador, Equador, Egypt, France, Finland, Germany, Great Britain, Georgia, Greece, Hong Kong, Honduras, Ireland, Israel, India, Italy, Iraq, Jordan, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Mexico, Morocco, Madagascar, Myanmar, Mauritius, Malaysia, Niger, Norway, New Zealand, Oman, Peru, Philippines, Pakistan, Poland, Potugal, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, Uzbekistan, South Africa and the Netherlands.