LICOX temperature probe, Model C8.B
FDA Recall Z-0623-05 · Posted 2005-03-15
Advertisement
Recalling FirmIntegra Neuro Sciences
Recall StatusTerminated
Product CodeGMW
Date Initiated2005-01-21
Date Terminated2005-07-13
Product Quantity19
Reason for Recall
LICOX Model C8.B temperature probes manufactured by GMS mbH, an Integra company, may exhibit progen (endotoxin) levels above the FDA limit of 2.15 endotoxin units per device. Elevated endotoxin levels may result in an increased patient risk of fever.
Firm Action
LICOX Model C8.B temperature probes and LICOX IM3.ST Complete Brain IMC Probe Kits manufactured by GMS mbH, an Integra company, may exhibit pyrogen (endotoxin) levels above the FDA limit of 2.15 endotoxin units per device. Elevated endotoxin levels may result in an increased patient risk of fever.
Root Cause
Other
Distribution Pattern
LA, CA, CO, DC, DE, FL, GA, IL, IN, LA, MA, MD, MI, MN, MT, NC, NE, NY, OH, OR, PA, SD, TN, TX, VA
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.