Completed

Reverse Shoulder System e+, REF 509-01-032, RSP Humeral Socket Insert, Sz. 32mm, Semi-Constrained, Sterile H2o2, CE 2797

FDA Recall Z-0625-2024 · Posted 2024-01-02

Recalling FirmEncore Medical, LP
Recall StatusCompleted
Product CodePHX
Date Initiated2023-10-05
Product Quantity17 units

Reason for Recall

Mislabeling

Firm Action

On October 5, 2022, DJO Global (Enovis company) issued a "Urgent Medical Device Recall" Notification via E-Mail. DJO asked consignees to take the following actions: 1. This notice needs to be passed on to all those within your organization or to any organization where the potentially affected product has been transferred. 2. Review your stock for the affected products and lot numbers for the items to be returned. 3. Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 to place a replacement order and receive an RMA number. You will be provided with a shipping label to return the affected product. 4. Return all affected devices (quantities listed above) using the RMA number. Report any adverse events related to this issue to Customer Service. Additionally, adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Root Cause

Packaging process control

Distribution Pattern

US Nationwide distribution in the states of VA, TN, MN, IN, RI, ME, TX, LA, AL, FL, CA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →