Recall StatusTerminated
Product CodeKOG
Date Initiated2010-02-19
Date Terminated2013-07-30
Product Quantity1,881 units (1514 in US)
Reason for Recall
The recall was initiated because Advanced Sterilization Products (ASP) received three customer
reports of elevated temperature of the high-level disinfectant solution used in the ASP AER as a result of a
possible temperature monitoring system malfunction. This recall communication advised consignees that if high-level
disinfectants are exposed to elevated temperatures, fumes may occur and resul
Firm Action
Advanced Sterilization Products sent an URGENT: PRODUCT CORRECTION letter dated February 19, 2010, to all customers. The letter identified the product, the problem, and the action to be taken by the customer. The recall communication was initiated with ASP forwarding a "Urgent: Product Correction" letter with attached customer fax-back response form to the affected customers.
Customers were also instructed to complete and sign the enclosed customer fax-back response form indicating that they have disconnected the heater on the ASP AER according to the instructions provided and fax the form to 1-888-345-1435.
If customers have any questions or need assistance, they were instructed to call ASP Professional Services at 888-783-7723, option 3 then option 1.
Distribution Pattern
Worldwide Distribution -- United States, Argentina Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador,
Hungary, Mexico, Panama, Guatemala, Honduras, Poland, Puerto Rico, Virgin Islands, Singapore, Taiwan,
Venezuela, Croatia, and Bulgaria.