Jarvik 2000 Ventricular Assist System external components: JHI-212 Y Cable, JHI-202 Lithium-ion Battery Cable, JHI-204 Extensions Cable, JHI-201 Pedestal Cable

FDA Recall Z-0629-2019 · Posted None

Recalling FirmJarvik Heart Inc
Recall StatusTerminated
Product CodeDSQ
Date Initiated2018-10-15
Date Terminated2020-05-01
Product Quantity274

Reason for Recall

There is a potential for the male connectors on external cables to become damaged. This could include a detached male connector housing, misaligned connector, cracked connector housing, and bent pins.

Firm Action

Urgent Field Safety Notice letters dated 10/5/18 were distributed to customers.

Root Cause

Device Design

Distribution Pattern

Products were distributed to the following US states: FL, NY, and UT. Products were distributed to the following foreign countries: Croatia, Czech Republic, France, Hungary, India, Italy, and Japan.