Stryker Color Cuff, REF 5921-034-135 Disposable Tourniquet Cuffs 34 in x 4 in (86 cm x 10 cm) Single Bladder, Single Port, QuickConnect Rx only, STERILE

FDA Recall Z-0630-2014 · Posted 2014-01-06

Recall StatusTerminated
Product CodeKCY
Date Initiated2013-12-06
Date Terminated2014-02-27
Product Quantity5 boxes (50 units)

Reason for Recall

The Color Cuff Sterile Disposable Tourniquet Cuffs with Quick-Connect Fitting may have an incorrect end-connector. A customer reported having received a luer connector instead of a quick connect fitting. An incorrect connector may lead to an inoperable system due to incompatibility with the Smart Pump.

Firm Action

Distributor was notified by E-mail 11/26/2013, with follow-up UPS letters sent to Hospital Risk/ Materials Managers identifying the product, reason for the recall, and risk. The notice asked consignees to find and quarantine all implicated product, with hospitals requested to complete the Business Reply Form and FAX to 866-521-2762. Stryker will email a pre-paid shipper and credit all accounts. Question and concerns should be addressed to Kelly Jo Whipple 269-389-2921 or [email protected]

Root Cause

Component change control

Distribution Pattern

US Distribution: including states of: MI, OH, and TX.