VARiS 1.4G and Vision/SomaVision Radiation Therapy Systems; VARiS versions 6.1.33 and 6.2.19 in conjunction with Eclipse 6.x/Soma Vision 6.x

FDA Recall Z-0633-03 · Posted 2003-03-12

Recall StatusTerminated
Product CodeIYE
Date Initiated2003-01-31
Date Terminated2003-09-17
Product Quantity12 units

Reason for Recall

The product has a software problem. This could result in a treatment without a motorized wedge where one had been planned.

Firm Action

On 1/16/03 and 2/4/03 (2nd notification) the firm initiated the recall and their notification was via letters to inform its consignees about the new VARiS version 6.2.27 release.

Root Cause

Software design

Distribution Pattern

Product was released for distribution to 7 hospitals in the US. Distribution was limited to the following States: CA, MA, NC, SC, PA, and LA. 1 (one) Canadian distribution includes Fraser Valley Cancer Center, Dept. of Radiation Oncology, 13750 96th Ave., Surrey, BC, V3V 1Z2, Canada. Foreign distribution includes Austria, Germany and Finland. The recall was appropriately extended to the user level; i.e., the hospitals/physicians who received the recalled product.