Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

KWIK-STIK(TM) labeled as containing Actinomyces odontolyticus. Each KWIK-STIK" unit contains a lyophilized pellet of a single microorganism strain, a reservoir of hydrating fluid and an inoculating swab. Each device is sealed within a laminated pouch that contains a desiccant to prevent adverse moisture accumulation. A KWIK-STIK" catalog number ending in P is a duopak and contains two KWIKSTIK" sticks.

FDA Recall Z-0636-2022 · Posted None

Recalling FirmMicrobiologics Inc
Recall StatusTerminated
Product CodeJTR
Date Initiated2017-03-02
Date Terminated2022-04-22
Product Quantity2 units

Reason for Recall

Potential for product labeled as Actinomyces odontolyticus is actually Eggerthella lenta.

Firm Action

Microbiologics notified customers on about 03/02/2017, via letter titled "Medical Device Recall." Customers were instructed to: 1. DO NOT USE any lot number 939-47-1 or 939-47-2 you may have in inventory. 2. QUARANTINE or DISCARD the product to prevent further use. 3. RETURN any remaining inventory you have of this product to Microbiologics via FED-EX though the use of FED-EX account number 355830926, and mark box Attention: Technical Manager. 4. Acknowledge your receipt of this information by completing the attached form and returning to Microbiologics by e-mailing to [email protected] or by faxing to 320.253.6250.

Root Cause

Labeling False and Misleading

Distribution Pattern

International distribution in the countries of Canada and Germany.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑