Terminated

LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E.

FDA Recall Z-0637-05 · Posted 2005-03-18

Recalling FirmDiasorin Inc.
Recall StatusTerminated
Product CodeMRG
Date Initiated2005-02-14
Date Terminated2006-07-22
Product Quantity131

Reason for Recall

Results of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator values.

Firm Action

DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.

Root Cause

Other

Distribution Pattern

MI, NJ, Germany, Spain and Belgium

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →