Liaison 25 OH Vitamin D Assay, Manufacturer part #310900, Lot #114668
FDA Recall Z-0638-05 · Posted 2005-03-18
Advertisement
Recalling FirmDiasorin Inc.
Recall StatusTerminated
Product CodeMRG
Date Initiated2005-02-14
Date Terminated2006-07-22
Product Quantity31 kits
Reason for Recall
An internal inspection indicates that Calibrator 2 and Conjugate are switched in their positions in the integral. This will result in Calibrator 1 RLU's similar to that of background RLU, and Calibrator 2 RLU's near expected range. A curve can not be calculated from this data.
Firm Action
DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.
Root Cause
Other
Distribution Pattern
MA, Spain, Germany & Belgium