ROSA Brain 3.0 Stereotaxic instrument Computer-assisted surgical device. Intended for the spatial positioning and orientation of instrument holders to be used by neurosurgeons.

FDA Recall Z-0638-2018 · Posted None

Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeHAW
Date Initiated2017-05-01
Date Terminated2020-07-13
Product Quantity16

Reason for Recall

Robot arm being sent to the wrong position

Firm Action

Urgent Medical Device Recall (Correction) notices were sent to customers. The letter provided to hospital risk managers and surgeons identified the issue and their responsibilities. These responsibilities included: - Stop using the micromovement function in isocentric mode during Brain Surgery with ROSA Brain 3.0 device. -Reviewing the notification to ensure comprehension of the steps required to be taken prior to the permanent correction that will be implemented on site by a service engineer. - Completing and returning Attachment 1 (Certificate of Acknowledgement) to the e-mail address provided. The firm has prepared a software correction that will be deployed and implemented on site by a MEDTECH representative.

Root Cause

Software design

Distribution Pattern

Distributed to CA, DC, FL, IL, KY, MA, MN, NC, NY, OH, and PA.