Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System

FDA Recall Z-0639-2024 · Posted 2024-01-04

Recall StatusOpen, Classified
Product CodeNKB
Date Initiated2023-10-16
Product Quantity3

Reason for Recall

Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.

Firm Action

On 10/16/23, recall notices were sent to customers asking them to do the following: 1) Locate, quarantine, and return affected screws to the recalling firm. 2) Complete and return the inventory return certificate form via email to [email protected] Recall-related questions or concerns can be directed to please email [email protected].

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the states of MN, NE.