Liko Mobile Lifts: Primarily intended for use in nursing homes LikoM230 Model: 2050015

FDA Recall Z-0640-2022 · Posted None

Recalling FirmHill-Rom, Inc.
Recall StatusOpen, Classified
Product CodeFSA
Date Initiated2021-12-23
Product Quantity1215 units

Reason for Recall

Missing bushing on Liko M220 and M230 causing wear and potential for patient falling

Firm Action

HillRom issued Urgent Medical Device Correction Letter (FSCA Identifier: FA 2021-010-002-LUL-001) via third party on 23 December 2021 via UPS Ground Service to Distributors and Direct Users. Letter states reason for recall, health risk and action to take: 1. Review the product label on your Liko M220/M230 to determine if it is impacted by this field action by identifying the manufacturing date of the device (i.e. Manufacturing date between 2016-DEC-27 and 2021 SEP-01). See figure 2. 2. Please periodically inspect your Liko device for damage or wear. Report visible wearing or damage to your Technical Support Team do not use the device until Technical Support can inspect and repair where necessary. 3. Please fill out the attached response form and return it to Hillrom [email protected] within two weeks Please share this Field Safety Notice with your end users and complete the attached response form and return to [email protected] within two weeks. Contact [email protected] to receive an electronic copy of this notification and response form for onward distribution. Action to be taken by the Hillrom: Upon receipt of the response form, identifying potentially impacted devices, Hillrom or an official Hillrom distributor/representative will be in contact to schedule inspection of your device to confirm presence of the bushing. Where it is confirmed the device is missing the bushing, a replacement slingbar with the bushing will be installed. If you have any questions regarding this safety notice, please contact Hillrom Technical Support at 800-445-3720 or [email protected]

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Austria, Austria, Bahrain, Canada, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Norway, Saudi Arabia, Sweden, Slovenia, Taiwan (Province of China), South Africa, Oman.