Catalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser System (International) - Catalys I 200-240V ~, 50/60 Hz, 15A, (2) 15A Resettable CB: Tested and complies with FCC Part 15 Class A. OptiMedica 1310 Moffett Park Drive Sunnyvale, CA 94089 The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.

FDA Recall Z-0649-2014 · Posted 2014-01-08

Recall StatusTerminated
Product CodeOOE
Date Initiated2013-12-20
Date Terminated2014-06-26
Product Quantity78 (39 in US, 39 internationally)

Reason for Recall

Potential for unexpected heating of the transformer mounting bolt located within the system chassis and the subsequent melting of a non-slip computer shelf mat, which may manifest an odor and visible smoke.

Firm Action

The firm, OptiMedica, issued an "ADVISORY NOTICE" dated December 20, 2013 to all customers with a Catalys Precision Laser System via Federal Express. OptiMedica also released a Field Service Bulletin (Document No: SB-00065) internally for the impacted product with instructions to implement the corrections through the OptiMedica service organization. The notice describes the product, problem and actions to be taken. The customers were instructed to complete the site header information on the attached fax-back form and return via fax to 1-408-890-4681 within the next three (3) business days or mail to OptiMedica at 1310 Moffette Park Drive, Sunnyvale, CA 94089. If you have any questions related to the operation of the Catalys Precision Laser System or the documentation requirements associated with this letter, please call OptiMedica Technical Support at 1-855-662-0911 and you will be directed to an OptiMedica representative who will be pleased to assist you.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution: US (nationwide) including states of: TX, NC, OH, IA, UT, KY, ME, CA, NY, MA. NJ, IL, PA,, TN, FL, MO, GA, HI, SC, NE, MI, and GA and countries of: Dominican Republic, Germany, Switzerland, Australia, Spain, Japan, United Kingdom, Canada, France, The Netherlands, Austria, and Singapore.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑