Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Veradius Unity, Model #718132

FDA Recall Z-0649-2022 · Posted None

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2021-12-02
Product Quantity12 US; 44 ROW

Reason for Recall

Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.

Firm Action

On December 2, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. The Medical Device Correction includes the following instructions for the Customer: o Ensure that the wired foot switch is always connected to the system. o Immediately start using the wired foot switch or hand switch in the event the wireless foot switch loses connection. o Perform a daily check for damage and correct functioning of the hand switch and wired foot switch as described by the User Routine Checks section in your Instruction for Use. o Circulate this notice to all users of this device so that they are aware of the issue./ o Place this Medical Device Correction with the documentation of the Philips Zenition 50, Zenition 70 or Veradius Unity system. o Return the attached reply form to Philips to confirm that: 1. The Zenition 50, Zenition 70 and Veradius Unity systems wired foot switch is connected to the system 2. The users of the system have reviewed and understood this Medical Device Correction. If you need further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377.

Root Cause

Under Investigation by firm

Distribution Pattern

Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑