Stryker Medical Stair-PRO Evacuation Chair civilian use stair chair, patient transport device, Model 6253.
FDA Recall Z-0650-2009 · Posted 2009-02-04
Advertisement
Recalling FirmStryker Medical Div. of Stryker Corporation
Recall StatusTerminated
Product CodeFPP
Date Initiated2008-11-18
Date Terminated2010-10-27
Product Quantity102
Reason for Recall
Handle could come off. The bolts may back out over time, resulting in a risk of dropping the patient and/or injury to transport personnel, a delay in medical treatment, or a delay in evacuation.
Firm Action
Consignees were notified by Urgent Medical Device Correction letter dated 11/18/08 and given the serial numbers of their units, an upgrade kit and instructions for performing replacement of the bolts. Consignees were informed Stryker will perform the upgrade if requested.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution: USA, India, Netherlands and United Kingdom.