Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood.

FDA Recall Z-0650-2014 · Posted 2014-01-09

Recall StatusTerminated
Product CodeCGA
Date Initiated2013-10-10
Date Terminated2014-03-24
Product Quantityest. 200,000

Reason for Recall

Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage

Firm Action

On November 18, 2013, an URGENT Field Correction Letter was sent to all StatStrip Glucose Hospital meter, StatStrip Lactate Hospital Meter and StatSensor Creatinine Hospital Meter customers via email to the Point of Care Coordinator to alert users that lithium batteries that are utilized beyond their intended useful life may swell and/or crack posing a potential safety hazard. In some cases, the swelling may potentially cause the battery door on the meter to dislodge and render the meter inoperable.

Root Cause

Component design/selection

Distribution Pattern

Nationwide Distribution.