CryoPatch SG Pulmonary Human Cardiac Patch. The patch is cryopreserved in a tissue culture medium containing cryoprotectants within the innermost pouch of a three-pouch packaging system. The packaging system not only withstands ultracold temperatures, but also allows for aseptic introduction of the patch into the operating environment. Supercooling by liquid nitrogen boost is begun prior to crystallization to minimize ice crystal damage to the patch matrix. Finally, the patch is transferred for long-term storage at or below 135 C. CryoPatch SG is distributed in three anatomic configurations: pulmonary hemi-artery, pulmonary trunk and pulmonary branch.

FDA Recall Z-0653-2021 · Posted None

Recalling FirmCryoLife, Inc.
Recall StatusTerminated
Product CodeDXZ
Date Initiated2020-11-10
Date Terminated2021-10-26
Product Quantity2 units

Reason for Recall

On November 10, 2020, it was discovered that there are post- processing microbial culture test failures for other allografts, processed using a TRIS saline solution lot that was also used in processing allografts subjected to the recall.

Firm Action

A Recall notification letter titled "Urgent - Voluntary Human Tissue Recall" was sent to consignees via email or hand delivered by firm representatives. On 11-Nov-2020, 4 letters were delivered to consignees. On 12-Nov-2020, 2 letters were delivered to consignees. The letter instructs the consignee to complete and return the enclosed postcard affirming that they received the notification. The affected product should be quarantined and returned to CryoLife immediately, if it has not been implanted. CryoLife Field Representatives were to provide information necessary to arrange for the return of the affected product. If the product has been implanted and the consignee has questions or requires clarification, the consignee is to contact Dr. Gregory Ray, CryloLife Medical Director at 800-438-8285.

Root Cause

Material/Component Contamination

Distribution Pattern

Distributed nationwide to GA, AZ, TX, MO, and OK.