Complete SE Biliary Stent System, SC5120L (5 mm x 12 mm), lot number 0000983799, UPC code 00613994186478, use by date 03/17/2011, manufactured by Medtronic CardioVascular, Santa Rosa, CA. Indicated for use in the palliation of malignant neoplasms in the biliary tree.
FDA Recall Z-0654-2010 · Posted 2010-01-19
Recalling FirmMedtronic CardioVascular
Recall StatusTerminated
Product CodeFGE
Date Initiated2009-07-17
Date Terminated2010-06-21
Product Quantity20 units
Reason for Recall
Mislabeling-- A lot of 5 mm x 120 mm stents were labeled as 6 mm x 80 mm stents and vice versa.
Firm Action
Medtronic, Inc. issued an "Urgent - Voluntary Medical Device Recall" communication dated July 17, 2009 asking consignees to segregate labeled lot numbers until a representative from the firm arrived to retrieve the products.
For further information, contact your local Medtronic representative or call Medtronic, Inc. at 1-707-525-0111.
Root Cause
Labeling mix-ups
Distribution Pattern
Product was distributed to 24 consignees in TX, MO, PA, MN, NY, IL, CA, IN, AZ, TN, OH, MI, and CT.