Terumo Perfusion System 8000 Base, 5 pump, 115v; Model 8000; Catalog No. 16401.

FDA Recall Z-0656-2007 · Posted 2007-03-27

Recall StatusTerminated
Product CodeDTW
Date Initiated2007-01-12
Date Terminated2008-06-19
Product Quantity36

Reason for Recall

The hand crank may crack in the bearing/shaft area due to a manufacturing defect. (Note: Some of the hand cranks subject to this recall were shipped as replacements for hand cranks recalled in the Summer of 2006.)

Firm Action

Consignees were notified by phone between 1/12/07 and 2/23/07 and informed replacement hand cranks would be shipped.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Belgium, Hong Kong, Japan, Taiwan and United Arab Emirates.