Recall StatusTerminated
Date Initiated2019-11-07
Date Terminated2021-04-23
Product Quantity48,428
Reason for Recall
A potential exists that certain LIGAMAX devices within
the impacted lots may have small holes in the Tyvek lidding which could result in a breach of sterility.
Firm Action
On November 7, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating:
URGENT: MEDICAL DEVICE RECALL (REMOVAL)
LIGAMAX 5mm Endoscopic Multiple Clip Applier
Product Codes: EL5ML
(Multiple Lot Numbers)
Dear Valued Customer
Ethicon has initiated a voluntary medical device recall (removal) of specific lots of LIGAMAX 5mm Endoscopic Multiple Clip Applier. Ethicon identified that a potential exists that certain LIGAMAX devices within the impacted lots may have small holes in the Tyvek lidding which could result in a breach of sterility. In the event that the sterility barrier has been breached, there is a chance that a pathogen may be introduced and if unrecognized or untreated may cause life threatening infection. As the population at risk is most likely to receive prophylactic antibiotics the probability of harm is extremely rare. We have identified the root cause and have implemented corrective actions to address the issue and prevent reoccurrence. As of the date of this communication, no adverse events or complaints have been reported for this issue.
Our records indicate that you may have ordered or received product subject to this recall. PLEASE DISTRIBUTE THIS INFORMATION TO ALL STAFF WITHIN YOUR FACILITY WHO USE LIGAMAX 5mm ENDOSCOPIC MULTIPLE CLIP APPLIER. The dates of distribution for affected products were from July 31, 2019 thru October 4, 2019.
EFFECTIVE IMMEDIATELY DO NOT USE OR DISTRIBUTE THE FOLLOWING PRODUCT CODE/ LOT:
Product Name: LIGAMAX
Product Code: EL5ML
Product Lots: T93X4R, T93X9D, T93Y2Z, T93Y70, T93Y96, T93Z52, T93Z76, T9401C, T9408F
T94118, T9416M, T9422K, T9426Z, T94402, T94536, T94567, T9461M, T94667, T94721, T94758, T94807, T9484Y, T9487J, T9491A, T94A9K, T94C2Z, T94C8N, T94D2N, T94E66, T94F3C, T94F7F, T94G99, T94H1N
T94H78
Description/Size: 5mm Endoscopic Multiple Clip Applier
Note: This recall is specific to lots described.
Health care practitioners who have treated pat
Root Cause
Under Investigation by firm
Distribution Pattern
US: Nationwide
International: ARGENTINA ARUBA AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CHINA COLOMBIA DENMARK ESTONIA FINLAND FRANCE GERMANY INDIA INDONESIA IRELAND JAPAN KOREA KUWAIT MALAYSIA MEXICO NETHERLANDS NEW CALEDONIA NEW ZEALAND PANAMA
PHILIPPINES PUERTO RICO QATAR SAUDI ARABIA SINGAPORE SLOVENIA SWEDEN SWITZERLAND TAIWAN UK
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.