ARROWg+ard Blue Two Lumen Hemodialysis Catheterization Kit for High Volume Infusions; Product Codes: AK-22142-CDC, AK-22142-F, AK-25142-CDC, AK-25142-F; Exp. Dates Feb 2018 - Nov 2018
FDA Recall Z-0658-2018 · Posted 2018-02-20
Advertisement
Recalling FirmArrow International Inc
Recall StatusTerminated
Product CodeMPB
Date Initiated2017-10-17
Date Terminated2020-04-14
Product Quantity5,926 total
Reason for Recall
These finished good kits may contain the incorrect Springwire Guide (SWG).
Firm Action
On October 17, 2017 an Urgent Medical Device recall letter from Arrow International was issued to customers requesting that customers review their inventory for product, which should immediately be discontinued and quarantined. The products are then to be returned to the firm. Questions or concerns can be directed to: Customer Service at 18662466990.
Root Cause
Process control
Distribution Pattern
Distribution US Nationwide