5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart (TAH-t). Part number 595000-001
FDA Recall Z-0659-2017 · Posted None
Advertisement
Recalling FirmSynCardia Systems Inc.
Recall StatusTerminated
Product CodeLOZ
Date Initiated2016-10-21
Date Terminated2017-02-21
Product Quantity5
Reason for Recall
The Main Printed Circuit Board Assembly (PCBA) of the affected Freedom Drivers may fail and cause the Freedom Driver to stop functioning without visual or audible alarms, resulting in the loss of life-sustaining function.
Firm Action
On 10/21/16 customers were notified by phone. Customers were informed that patients who were being supported by affected Drivers must be switched to unaffected Drivers.
The firm will provide unaffected Freedom Drivers to replace the affected Drivers. Affected Drivers must be return to the firm. Instructions for return provided, and Acknowledgment Form must be completed.
Contact Information For Questions
Mary Pat Sloan
Senior Vice President
Global Certification and Logistics
SynCardia Systems, LLC
1992 E Silverlake Rd
Tucson, AZ 85713
Direct 520-547-7470
Toll-free 866-480-1122 Ext 1308
Monday-Friday 7:30 AM to 5:00 PM Pacific/Arizona time
Hotline 24/7 866-771-9437
www.syncardia.com
On 10/27/16 customer notification letters were sent.
The letter provided instructions on how to recognize the device could fail.
The patient and caregiver will recognize that an affected Driver is failing because the Freedom Driver will stop pumping without generating any sound or any visual or audible alarms.
When functioning, the Freedom Driver exhibits a distinctive pumping sound; when it stops pumping the lack of sound is noticeable.
Generally, if the Freedom Driver function is interrupted, a visual and/or audible alarm will occur. However, as a result of this component failure, when affected Freedom Drivers stop functioning, there will be no visual or audible alarms, and no pumping sound.
Actions to Be Taken by The Customer/User
SynCardia has provided unaffected Freedom Drivers to the two hospitals to replace the affected 5000 series Drivers. All patients with affected Freedom Drivers have undergone the exchange and received unaffected Drivers.
The hospitals have returned all the affected Freedom Drivers to SynCardia.
The hospitals are requested to complete the Recall Acknowledgement and Receipt Form below and return to SynCardia.
Root Cause
Process control
Distribution Pattern
US distribution to Virginia and Arizona.