AC Power Cords, 110V Manufactured by Electri-Cord Manufacturing Company in Westfield, PA. Power Cord #M4268840 used with First Step All-In-One Canadian French Models 227500-09 and 227500, First Step All-In-One English Model 227500, Therakair Visio Models M8259978 and M6257763 pump only, and Proficare - Canada Model M8259988
FDA Recall Z-0664-2010 · Posted 2010-01-25
Advertisement
Recalling FirmKCI USA, Inc.
Recall StatusTerminated
Product CodeIKY
Date Initiated2009-11-04
Date Terminated2011-01-21
Product Quantity1,384
Reason for Recall
Power cords may crack and fail inside plug with potential for fire hazard.
Firm Action
Kinetics Concepts, Inc. (KCI) issued an "Urgent Medical Device Correction" letter dated November 4, 2009. Consignees were informed of the affected product and instructed to take necessary action to avoid any further issues associated with the product.
For further information, contact KCI's technical support team at 1-800-275-4524 option #3.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide and Canada
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.