MICRO-X Rover Mobile X-ray System, # MXU-RV35
FDA Recall Z-0664-2025 · Posted 2024-12-12
Advertisement
Recalling FirmMicro-X Ltd.
Recall StatusOpen, Classified
Product CodeIZL
Date Initiated2023-05-31
Product Quantity5
Reason for Recall
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
Firm Action
The firm will notify consignees using a service bulletin that was already posted. Additionally, the firm will implement a firmware update to extend the timer (4 ms to 10 ms minimum) to ensure even short exposures achieve the necessary image quality for diagnostic use.
Root Cause
Software design
Distribution Pattern
Worldwide and US Nationwide Distribution: NJ, PR, CA