MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
FDA Recall Z-0665-2025 · Posted 2024-12-12
Advertisement
Recalling FirmMicro-X Ltd.
Recall StatusOpen, Classified
Product CodeIZL
Date Initiated2023-11-28
Product Quantity18
Reason for Recall
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
Firm Action
On 11/28/23 Modification 17 was sent to customers telling them that Control Boards need to be replaced with Control Board PCBA, 12425-04 that will eliminate the issue.
For technical assistance contact: [email protected], 1-206-249-8764
Firm asked distributors to do the following: On next visit please confirm that the bulletins were completed and return the documentation.
Root Cause
Device Design
Distribution Pattern
US Nationwide Distribution: NJ, CA, PR