Terminated

AC Power Cords, 110V Manufactured by Electri-Cord Manufacturing Company in Westfield, PA. Power Cord #M4268882 used with InfoVAC Model M825999,, InvoVac Canadian Model M8260027, and Vac ATS Model 8259968.

FDA Recall Z-0666-2010 · Posted 2010-01-25

Recalling FirmKCI USA, Inc.
Recall StatusTerminated
Product CodeOMP
Date Initiated2009-11-04
Date Terminated2011-01-21
Product Quantity24,640

Reason for Recall

Power cords may crack and fail inside plug with potential for fire hazard.

Firm Action

Kinetics Concepts, Inc. (KCI) issued an "Urgent Medical Device Correction" letter dated November 4, 2009. Consignees were informed of the affected product and instructed to take necessary action to avoid any further issues associated with the product. For further information, contact KCI's technical support team at 1-800-275-4524 option #3.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide and Canada