Terminated

Terumo Advanced Perfusion System 1 (with 6 inch diameter roller pump); Catalog number 801041.

FDA Recall Z-0667-05 · Posted 2005-03-31

Recall StatusTerminated
Product CodeDTQ
Date Initiated2005-03-17
Date Terminated2005-09-16

Reason for Recall

The pumps will stop, show an overspeed error message, and will not restart using routine procedures if the user reduces the pump speed to zero RPM while the pump is operating in pulse mode.

Firm Action

Recall letters dated 3/17/05 were sent to customers advising them of the problem, the temporary fix, and that software will be upgraded.

Root Cause

Other

Distribution Pattern

Nationwide, Australia, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malasia, Mexico, Pakistan, South Africa, South Korea, Taiwan, Thailand and United Arab Emirates.