AC Power Cords, 110V Manufactured by Electri-Cord Manufacturing Company in Westfield, PA. Power Cord M426884 used with VAC ATS Canadian Model M8259994.
FDA Recall Z-0667-2010 · Posted 2010-01-25
Advertisement
Recalling FirmKCI USA, Inc.
Recall StatusTerminated
Product CodeOMP
Date Initiated2009-11-04
Date Terminated2011-01-21
Product Quantity649
Reason for Recall
Power cords may crack and fail inside plug with potential for fire hazard.
Firm Action
Kinetics Concepts, Inc. (KCI) issued an "Urgent Medical Device Correction" letter dated November 4, 2009. Consignees were informed of the affected product and instructed to take necessary action to avoid any further issues associated with the product.
For further information, contact KCI's technical support team at 1-800-275-4524 option #3.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide and Canada
Official Source