Terminated

Stryker GMRS Proximal Femoral Surgical Protocol; Global Modular Replacement System Proximal Femoral Resection for Large Segmental Replacements, Femur Prosthesis Surgical Literature, Stryker Orthopaedics, Mahwah, NJ 07430

FDA Recall Z-0668-2008 · Posted 2008-04-22

Recall StatusTerminated
Product CodeJDI
Date Initiated2007-12-18
Date Terminated2008-10-15
Product Quantity5869

Reason for Recall

The GMRS Proximal Femoral Surgical Protocol states " The GMRS Proximal Femoral components are fully compatabile with all Stryker V40 femoral heads." This statement is incorrect and conflicts with the Alumina Head label and packaging insert.

Firm Action

Stryker sent Urgent Product Recall notification letters and Product Accountability forms via FedEx on 12/21/2007, return receipt requested. An Urgent Product Recall letter dated 1/7/2008, was sent to surgeons who are GMRS users, advising that the GMRS Proximal femoral components are not compatible with Alumina ceramic heads.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution - USA, Canada, Mexico, China, Australia, Hong Kong, Singapore, Sweden, Germany, Brussels, the Netherlands, Italy, Poland, Dubai, United Kingdom, Spain, and South Africa.