LH 700 Series Hematology Analyzers. Part numbers 66056032, 6605633, 6605645
FDA Recall Z-0669-03 · Posted 2003-03-21
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeGKZ
Date Initiated2002-12-27
Date Terminated2003-07-26
Product Quantity234
Reason for Recall
Incorrect Hemoglobin result can be reported with software version 2A and higher.
Firm Action
Customers were contacted by telephone and mail from 1/27/03 to 2/3/2003. Notice included modified operating instructions. A software fix is being developed and will be installed by Beckman personnel.
Root Cause
Other
Distribution Pattern
Nationwide and Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.