LH 700 Series Hematology Analyzers. Part numbers 66056032, 6605633, 6605645

FDA Recall Z-0669-03 · Posted 2003-03-21

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeGKZ
Date Initiated2002-12-27
Date Terminated2003-07-26
Product Quantity234

Reason for Recall

Incorrect Hemoglobin result can be reported with software version 2A and higher.

Firm Action

Customers were contacted by telephone and mail from 1/27/03 to 2/3/2003. Notice included modified operating instructions. A software fix is being developed and will be installed by Beckman personnel.

Root Cause

Other

Distribution Pattern

Nationwide and Canada.