MISAGO RX Self Expanding Peripheral Stem
FDA Recall Z-0669-2017 · Posted None
Advertisement
Recalling FirmTerumo Medical Corp
Recall StatusTerminated
Product CodeNIP
Date Initiated2016-08-04
Date Terminated2017-10-19
Product Quantity7,965 units
Reason for Recall
During recent routine testing, the MISAGO Peripheral Stent was found not to conform to product specification related to stent diameter at the proximal/distal ends and/or the shape of the stent.
Firm Action
Terumo Medical Corporation issued an Urgent Product Recall Bulletin dated August 4, 2016, via Fed Ex.
Customers must immediately discontinue use and quarantine any available stock of the Misago peripheral stent to prevent further use.
The customers should complete the enclosed Inventory Confirmation form as soon as possible and email or fax the completed form
to [email protected] or (734) 330-2034, Attn: MISAGO Recall.
Customers are encourage you to contact Terumo Medical Corporation with any questions or concerns
-Terumo Medical Corporation Quality Department: 1.800.283.7866, Press 4 for list of departments, Press 7 for Quality Assurance; Monday Friday, 8 a.m. 5 p.m. ET
- Email Quality Assurance: [email protected]
- Customer Care: 1.800.888.3786
- Contact your local Terumo representative
Root Cause
Device Design
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.