Terminated

Intuitive brand Da Vinci S Surgical System, an Endoscopic Instrument Control System, Model SS2000 Patient Side Cart, Model PS2000, Part Number: 380267-01, IS2000, Top Level, 4-Arm; Distributed and Manufactured by: Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA

FDA Recall Z-0670-2007 · Posted 2007-03-29

Recall StatusTerminated
Product CodeNAY
Date Initiated2006-04-12
Date Terminated2007-12-05
Product Quantity24 units

Reason for Recall

Da Vinci S Surgical System's spinal pin could limit mechanical motion of the arm and render the system unavailable for surgical use, and may result in system failure and the need to convert to an alternative surgical technique.

Firm Action

The firm did not release notification to its consignees. The reworks were arranged as part of routine service visits initiated on 4/12/06.

Root Cause

Other

Distribution Pattern

Nationwide.