Terminated

NeoSpan SuperElastic Compression Staple w/Instruments, Size 10x10x10, Part Number T50 SN010

FDA Recall Z-0670-2019 · Posted None

Recalling FirmIn2bones USA, LLC
Recall StatusTerminated
Product CodeJDR
Date Initiated2018-11-14
Date Terminated2019-05-08
Product Quantity30 units

Reason for Recall

The NeoSpan Compression Staple (10x10x10) kits from two lots contain the wrong size drill bit. They contain a 3mm drill bit instead of a 2mm drill bit.

Firm Action

The firm initiated the recall by letter on 11/14/2018. The letter requested the return of the product.

Root Cause

Employee error

Distribution Pattern

TX, VA, PA, MO, NJ, FL, TN, NC, OH, WI, LA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →