AIA-PACK B12 (Vitamin B12) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of vitamin B12 (B12) in human serum on Tosoh AIA System analyzers.

FDA Recall Z-0671-2019 · Posted None

Recalling FirmTosoh Bioscience Inc
Recall StatusTerminated
Product CodeCDD
Date Initiated2018-11-30
Date Terminated2021-01-14
Product Quantity29645 units

Reason for Recall

Potential for fluorescein interference to cause inaccurately elevated results for certain Tosoh AIA assays

Firm Action

On November 30, 2018, Tosoh Bioscience issued Urgent Medical Device Recall notices to customers. Customers are advised to take the following actions: " Do not use the assays identified in notice for those patients who have recently had a fluorescein angiography procedure. For these patients, use an alternative testing method that does not use a fluorescein-based technology. " Inform medical professionals that the Tosoh AIA assays in the notice utilize a fluorescein-based technology and should not be used to test patient samples for patients who have recently undergone fluorescein angiography. " Continue to use the Tosoh AIA assays listed in notice only for patients who have not undergone a fluorescein angiography procedure exam, or on patient samples which have been taken prior to receiving the fluorescein angiography procedure. " Complete & return the Urgent Medical Device Recall Acknowledgement Form within 15-days of receiving the notification. " Maintain a copy of the notification with your laboratory records and forward this information to others who may have received this product. Should you have any questions, please contact Bernadette OConnell at 1(800) 248-6764 or by email at [email protected]. She will be able to answer any questions Monday-Friday from 9 AM to 5 PM (PST).

Root Cause

Device Design

Distribution Pattern

US Nationwide