Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the ENA 6 Screen ELISA kit assay.

FDA Recall Z-0672-05 · Posted 2005-04-06

Recalling FirmDiasorin Inc.
Recall StatusTerminated
Product CodeLLL
Date Initiated2004-04-20
Date Terminated2006-07-22
Product Quantity29

Reason for Recall

An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing.

Firm Action

Customers were sent a notification letter informing customers of an error in the Bio-Tek ELx800 programming specifications. Customers were asked to repeat testing for historical runs for RSLT (S/CO) that read from 0.95 to 1.0. A response form is asked to be faxed back to DiaSorin.

Root Cause

Other

Distribution Pattern

CT, FL, NY, WA, TX, AR, TN, CANADA