Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the ENA 6 Screen ELISA kit assay.
FDA Recall Z-0672-05 · Posted 2005-04-06
Advertisement
Recalling FirmDiasorin Inc.
Recall StatusTerminated
Product CodeLLL
Date Initiated2004-04-20
Date Terminated2006-07-22
Product Quantity29
Reason for Recall
An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing.
Firm Action
Customers were sent a notification letter informing customers of an error in the Bio-Tek ELx800 programming specifications. Customers were asked to repeat testing for historical runs for RSLT (S/CO) that read from 0.95 to 1.0. A response form is asked to be faxed back to DiaSorin.
Root Cause
Other
Distribution Pattern
CT, FL, NY, WA, TX, AR, TN, CANADA